Reports are printed incorrectly, Data cannot be transmitted to a computer – HEIDENHAIN ND 1400 User Manual
Page 162

148
QC-300 Series User’s Guide
Reports are printed incorrectly
Possible cause Possible solutions
Printed report setup is incorrect 1 Correct the report setup in the Print
setup screen
Wrong control characters are specified 1 Correct the control character configurations
in the Print setup screen
Data cannot be transmitted to a computer
Possible cause Possible solutions
Wrong serial cable is connected 1 Connect the computer to the QC-330 using an
RS-232 serial cable that does not include
crossed transmit and receive wires. The
correct cable can be ordered from Metronics
by specifying part number 11B12176
Wrong RS-232 port settings are specified
1 Change the RS-232 port settings to match the
computer’s serial port settings in the Port setup
screen
Wrong control characters are specified
1 Correct the control characters to reflect the
requirements of the receiving software in
the Print setup screen