Warning – Amprobe GP-2 Geo-Test User Manual

Page 10

Advertising
background image

GP-2 GeoTest

EN - 10

Press GO to take the measurement and read the result on the display. At the
end of the test, a screen similar to the following will be displayed.

NOTE!

If you keep pressing GO, the instrument takes more measurements
consecutively. When a new value is acquired, the symbol

blinks on the main

display, the instrument emits a short sound and the counter shown on the
secondary display on the left-hand side is updated. This counter indicates the
number of measurements calculated on the basis of the average resistance
value.

0.96

0.96

2

2 0.93

0.93

WARNING:

When the message “Measuring” appears on the display, the
instrument is measuring. Do not disconnect the alligator
clips during the measurement

.

Ex.:

if the operator takes three measurements consecutively, the instrument will
display:

- 1

st

measurement:

main display = measured resistance value (Ex: 0.90

)

secondary display on the left-hand side = 001 (no. of measurements = 1
means that 1 earth measurement has been taken)
secondary display on the right-hand side = average of the measurements
taken (in case just one measurement has been taken the average value is
equal to the measured value, in this case 0.90

)

- 2

nd

measurement:

main display = measured resistance value (Ex: 0.96

)

secondary display on the left-hand side = 002 (no. of measurements = 2
means that 2 earth measurements have been taken consecutively)
secondary display on the right-hand side = average of the measurements
taken ((Val1+Val2)/no. of measurements = (0.90+0.96)/2 = 0.93

)

- 3

rd

measurement:

main display = measured resistance value (Ex: 0.93

)

No. of earth resistance
measurements
included in the
calculation of the
average resistance

Average value of the resistance calculated
as:

(Val1+Val2+…+Valn)/(No. measurements)

Value of the
resistance

GO

Advertising