2... OMP-MODL System Base

Using the OMP-MODL

2-6

3-Button System Initialization:

A complete initialization of the logger that will clear data memory
and program memory can be performed using the ENABLE, STOP
and RESET buttons. This sequence is normally only used when a
unit is upgraded in the field with a new EPROM or in the event that
the Program memory has become corrupted due to unforeseen
events such as disassembly while powered up, improper insertion of
a PCMCIA card, exposure to an extreme noise noise glitch (for
example, from a nearby lightning strike) that has caused the
microprocessor to lose its place or lock-up or other malfunction.

To perform this 3-Button Initialization,

1. Depress and hold the ENABLE button

2. Momentarily, depress the STOP and RESET buttons

simultaneously.

3. After a second or so, release the ENABLE button.

4. Observe the Feedback LED. After a few seconds, the

Feedback LED should blink 10 times in succession. This
indicates that a complete system initialization has been
performed.
If the logger is equipped with a ML-DISP modules, after a
short sequence of display messages on the LCD, a
SYSTEM INITIALIZED message should display
momentarily indicating that the logger was properly
initialized. If this message does not display, repeat the
procedure.

After initialization, reprogram the logger with a new Net Program
and the unit is ready to operate.

RS-232 Serial Communications Port

A female 6/6 RJ-12 modular phone type jack is provided on the MLCPU-1
for RS-232 communications. A mating 6 conductor cable (CAR-4) plugs into
this port. The other end of the cable plugs into the 9-pin or 25 pin serial port
on a PC via a modular plug to DB-9F (P/N: RJDB-9H) or DB-25F (P/N:
RJDB-25H) adapter. Note that this port is not for direct connection of a
telephone line.