Teledyne 6600 User Manual

Page 142

Advertising
background image

1-52 Part III

Part III

Model 6600

Teledyne Analytical Instruments

c Check to see that lines are properly heat traced if
required for ambient temperature extremes and that the
sample will not boil from being too hot (for example:
(Maximum temperature allowed is:52 degrees C (limited
by the plastic filter housings), Internal sample system
operation Pressure limit is:150psig G for produced water).
The inlet sample pressure (457 psig) is reduced by a
forward pressure regulator (PR1) which is followed by a
protecting relief valve in the event of failure of PR1).

d Check and familiarize oneself of the analyzer and sample
system components and to assure the entire piping flow
agrees with the Piping Drawing

(see addendum

drawings)

1

Sample connections; Refer to Piping diagram

(see addendum drawings)

a

inlet-
-sample enters the pump.

b

returns
- sample flows through Cell then exits the
system to a common return

c

drain
Assure the common drain(s) is always free
flowing and always at atmospheric pressure
(unless back to process by suitable
differential or pumping) and will never
freeze or block (backup) under icy
conditions.

d

Grab Sample port: Refer to your particular
piping scheme for obtaining the sample for
analysis, etc.OBSERVE EXTREME
CAUTION
when operating any needle

Advertising
This manual is related to the following products: