Teledyne 6600 User Manual

Page 41

Advertising
background image

Oil in Water Analyzer

Operation 3

Part I 3-17

Teledyne Analytical Instruments

Note: Shut off the gas pressure before connecting it to the analyzer,

and be sure to limit pressure to 40 psig or less when turning it
back on.

Readjust the gas pressure into the analyzer until the flowrate through the

Sample Cell settles between 50 to 500 cc/min (approximately 0.1 to 0.4
SCFH).

Note: Always keep the calibration gas flow as close to the flowrate of

the sample gas as possible

3.4.1 Zero Cal

The Zero function on the main menu is used to enter the zero calibration

function. Zero calibration can be performed in either the automatic or manual
mode.

Make sure the zero fluid is flowing to the instrument. If you get a

CELL

CANNOT BE BALANCED message while zeroing skip to section 3.4.1.3.

3.4.1.1

Auto Mode Zeroing

Observe the precautions in sections 3.4 and 3.4.1, above. Enter the zero

function mode. The screen allows you to select whether the zero calibration is to
be performed automatically or manually. Use the

UP/DOWN switch to toggle

between

AUTO and MAN zero settling. Stop when AUTO appears, blinking, on

the display.

Select zero
mode: AUTO

Enter

to begin zeroing.

####.## ppm OIL
Slope=#.### C—Zero

The beginning zero level is shown in the upper left corner of the display. As

the zero reading settles, the screen displays and updates information on

Slope=

in percent/second (unless the Slope starts within the acceptable zero range and
does not need to settle further). The system first does a coarse zero, shown in
the lower right corner of the screen as

C—Zero, for 3 min, and then does a fine

zero, and displays

F—Zero, for 3 min.

Then, and whenever Slope is less than 0.01 for at least 12 sec, instead of

Slope you will see a countdown:

9 Left, 8 Left, and so fourth. These are

Advertising
This manual is related to the following products: