Part I: Control Section

Part I: 2-5

Teledyne Analytical Instruments

Table 2-2: Analog Range ID Output - Example

Range

Voltage (V)

Current (mA)

Range 1

0.25

8

Range 2

0.50

12

Range 3

0.75

16

Alarm Relays:
There are three alarm-circuit connectors on the alarm relays block

(under RELAY OUTPUTS) for making connections to internal alarm relay
contacts. Each provides a set of Form C contacts for each type of alarm.
Each has both normally open and normally closed contact connections. The
contact connections are indicated by diagrams on the rear panel. They are
capable of switching up to 3 ampers at 250 V AC into a resistive load
(Figure 2-6).

Figure 2-6: Types of Relay Contacts

The connectors are:

Threshold Alarm 1:

• Can be configured as high (actuates when
concentration is above threshold), or low
(actuates when concentration is below thresh old).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).

Normally open

Moving contact

Normally closed

Normally open
Moving contact