Oil in Water Analyzer

Operation 3

Part I 3-5

Teledyne Analytical Instruments

APPLICATION: Used to define the analysis ranges and application
(gas used).

MODEL: Displays model number and software version.

OUTPUT_CAL: 4-20 MA: Adjust 4 and 20 mA output.

The hierarchy of the system menu is shown in figure 3-1.

3.3.1 Setting up an AUTO-CAL

When proper automatic valving is connected, the Analyzer can cycle itself

through a sequence of steps that automatically zero and span the instrument.

Note: Before setting up an AUTO-CAL, be sure you understand the

Zero and Span functions as described in section 4.4, and
follow the precautions given there.

Note: If you require highly accurate AUTO-CAL timing, use external

AUTO-CAL control where possible. The internal clock in the
Model 6600 is accurate to 2-3 %. Accordingly, internally sched-
uled calibrations can vary 2-3 % per day.

To setup an

Auto–Cal cycle:

Choose System from the main manu. TheVFD will display five subfunc-

tions.

DIG_FILT SELF—TEST
PWD LOGOUT MORE

Select MORE and Enter

AUTOCAL HMGNZR

HOLD

CAL-HOLD-TIMER

MORE

Use

UP/DOWN to blink

AUTO—CAL

, and Enter. A new screen for

ZERO/SPAN set appears.

ZERO in Ød Øh off
SPAN in Ød Øh off

Use UP/DOWN to blink

ZERO

(or

SPAN

), then Enter again. (You won’t

be able to set

OFF to ON if a zero interval is entered.) A Span Every ...

(or

Zero Every ...) screen appears.

Zero schedule: OFF
Day: Ød Hour: Шh