Part III: 1-5

Oil in Water

Part III

Teledyne Analytical Instruments

B.

0-20ppm down to 0-10ppm oil ranges in high background

waters such as off-shore platforms, produced waters, sea water, wastewater,
effluents, ponds, bilge/deballasting treatments, on-board ship applications: (2
and 4
, if no sample pressure or continuous gravity feed available. (Note:
Assume sample inlet contains dissolved oil and is homogeneous).

C.

0-50ppm to 0-200ppm oil in high background waters such as

off-shore platforms, produced waters, sea water, wastewater, effluents, ponds,
bilge/deballasting treatments, on-board ship applications, tank farms, fuel
depots, rig-washing decks, etc., (1, 2 and 4 if no sample pressures or gravity
feed available). NOTE: for ranges higher than 200ppm oil a dilution system is
required.

Note: Assume sample inlet contains both dissolved and non-dissolved

oil with non-oil organic background compositions and is representative. It
should also be uniform and kept homogeneous up to the homogenization
step.

NOTE: By adjusting valve, V4 in a position for “F1 only”, filtering

versus “F1, F2” and F3 filtering during the auto-zero functioning selects
whether the customer wishes to measure “Total oil and grease recoverable”
or “non-dissolved oil” only. This becomes advantageous when environmen-
tal regulation agencies allow tolerable dissolved oil level compositions in the
waters. Many times, cost savings are realized in clean-up operations.

Oil in Water Piping Diagram (B71046)