Teledyne 6600 User Manual

Page 58

Advertising
background image

4 Maintenance

Model 6600 Oil in Water Analyzer

4-2: Part I

Teledyne Analytical Instruments

3.0 A MAX

Figure 4-1: Removing Fuse Block Cap and Fuse from Housing

2. Replace fuse by reversing process in step 1.

4.2

System Self Diagnostic Test

1. Press the System button to enter the system mode.

2. Use the < > arrow keys to move to More, and press Enter.

3. Use the < > arrow keys to move to Self-Test, and press Enter.

The following failure codes apply:

Table 5-1: Self Test Failure Codes

Power
0

OK

1

5 V Failure

2

15 V Failures

3

Both Failed

Analog
0

OK

1

DAC A (0–1 V Concentration)

2

DAC B (0–1 V Range ID)

3

Both Failed

Advertising
This manual is related to the following products: