background image

3 Operation

Model 6600

3-6 Part I

Teledyne Analytical Instruments

Use

UP/DOWN

UP/DOWN

UP/DOWN

UP/DOWN

UP/DOWN

to set the day interval, hour interval, then Enter

Enter to turn

ON the SPAN and/or ZERO cycles (to activate AUTO–CAL).

Use the UP/DOWN to toggle the field between ON and OFF. Press Enter to
return to The AUTO-CAL menu. You should be able to see that the screen has
been updated with your new input. Escape to return to the System menu.

For Oil & Water Samples only setting of the Zero schedule is need-

ed. Instrument will automatically perform a Span at the end of a sched-
uled Zero.

If instrument is turned off, the next time the instrument is powered, the

instrument will automatically perform a calibration cycle after 3 minutes of
entering the sample mode if AUTOCAL functions were on prior to shut down.

3.3.2 Password Protection

Before a unique password is assigned, the system assigns

TAI by default.

This password will be displayed automatically. The operator just uses the Enter
switch to be allowed total access to the instrument’s features.

If a password is assigned, then setting the following system parameters can

be done only after the password is entered: alarm setpoints, assigning a new
password, range/application selections, and curve algorithm linearization.
(APPLICATION and ALGORITHM are covered in the programming section.)
However, the instrument can still be used for analysis or for initiating a self-test
without entering the password. To defeat security the password must be
changed back to

TAI.

NOTE: If you use password security, it is advisable to keep a copy of

the password in a separate, safe location.

3.3.2.1

Entering the Password

To install a new password or change a previously installed password, you

must key in and ENTER the old password first. If the default password is in
effect, using ENTER three times will enter the default

TAI password for you.

Enter the System menu...

DIG_FILT AUTO—CAL
PWD LOGOUT MORE

Use the

UP/DOWN to scroll the blinking over to

PWD

, and Enter to select

the password function. Either the default

TAI password or AAA place holders for

an existing password will appear on screen depending on whether or not a
password has been previously installed.

Advertising