Teledyne 6600 User Manual

Page 56

Advertising
background image

3 Operation

Model 6600

3-32 Part I

Teledyne Analytical Instruments

the analyzer is zeroed and spanned as usual. Then, each special calibration gas,
for each of the intermediate calibration points, is flowed, in turn, through the
sensor. As each gas flows, the differential value for that intermediate calibration
point is entered from the front panel of the analyzer.

Note: The span gas used to span the analyzer must be >90% of the

range being analyzed.

Before starting linearization, perform a standard calibration. See section

4.4. To enter data:

From the System Functions screen—
1. Select

ALGORITHM , and Enter.

2. Select and Enter

SETUP.

3. Enter

AUTO from the Calibration Mode Select screen.

The Auto Linearize Mode data entry screen appears.

19.5 ppm SO2
Input(Ø) :20.00

5. Use the UP/DOWN switch to set the proper value of calibration

gas, and Enter. Repeat this step for each cal-point number as it
appears in the

Input (x) parentheses.

6. Repeat step 5 for each of the special calibration gases, from the

lowest to the highest concentrations. Escape when done.

To end the session, send:

st<enter>
st<enter>

to the analyzer from the computer.