2 Installation

Model 6600

2-4: Part II

Teledyne Analytical Instruments

The span fluid must be the major component of the sample mixed with

a small amount of the component of interest. The concentration must be 60 to
80% of the range or the widest range of the instrument (if the instrument provides
more than one range).

2.4.2 Calibration

Refer to Section 3.3.8 section I of the manual to determine how to

manipulate the mode setting. Two calibration methods are available.

1. Calibration with zero and span fluids.
2. Calibration with a span filter.

Method One:

1.

Inject zero fluid and set zero as referred in Part I

2.

Inject span fluid and set the concentration of the span fluid with
the span procedure referred in Part I

Method Two:

1.

Determine the span setting using Method One.

2.

Activate the span filter (as referred in section 3.3.8) Part I

3.

Record the display reading (this is the span filter reading and

must be recorded).
4.

You can calibrate the instrument now with the span filter.

Power up the system, and test it as follows:

1. Repeat the Self-Diagnostic Test.