Teledyne 6600 User Manual

Page 79

Advertising
background image

Oil in Water Analyzer

Maintenance 3

Part II 3-3

Teledyne Analytical Instruments

2.

Check the UV source.

NOTE: Be sure to wear UV filtering eye goggles.

3.

Check the solenoid valves.

3.5

Service Procedures and Adjustments

3.5.1 Electronics

TAI aligns the system’s electronics. However, you may

need to touch up the circuitry, using the following procedure.

Equipment Required:

Oscilloscope (dual trace is preferred, but not required) To observe

oscilloscope test points switch the vertical input selector of the scope to DC.

Switch to AC to observe the demodulator switch signals.

DVM (Digital Voltmeter)

3.5.2 Power Supply Test Points

Measure +15 volt ±1 volt DC and -15 volt ±1 volt DC on the differential

power supply PC board in the control unit. Refer to the power supply schematic
in the back of the manual to identify the power supply test points, or section 3.6
in this chapter.

3.5.3 Setup of the Signal Processing Front-End

Amplifiers

Fill the sample cell with air or a stable fluid, such that the photo

energy that strikes the detector is constant. A stable fluid is distilled or tap
water. This step may be omitted when the system is stable in its present
state.