Oil in Water Analyzer

Operation 3

Part I 3-25

Teledyne Analytical Instruments

3.6.1 Manual (Select/Define Range) Screen

The Manual range-switching mode allows you to select a single, fixed

analysis range. It then allows you to redefine the upper and lower limits, for the
range.

Enter the Range function to start the Range function.

Select range
mode: MANUAL

If above screen displays, use the

UP/DOWN switch to Select

MANUAL,

and Enter.

Select range to run
—> Ш1 Ш2 Ш3 <—

NOTE: Oil in Water applications require single range (01) measurement.

Use the

UP/DOWN switch to select the range: 01, 02, 03, or 04. Then

Enter

.

Gas use: OIL
Range: Ø — 100 ppm

The high-end of the range field should blink first. Use UP/DOWN switch

to change the value of the field, Enter to move to the low-end of the range field.

Escape

to return to the previous screen to select or define another range.

Enter

to return the to the Analyze function.

3.6.2 Auto Screen

Autoranging will automatically set to the application that has at least two

ranges setup with the same gases.

In the autoranging mode, the microprocessor automatically responds to

concentration changes by switching ranges for optimum readout sensitivity. If the
upper limit of the operating range is reached, the instrument automatically shifts
to the next higher range. If the concentration falls to below 85% of full scale of
the next lower range, the instrument switches to the lower range. A correspond-
ing shift in the DC concentration output, and in the range ID outputs, will be
noticed.

The autoranging feature can be overridden so that analog output stays on a

fixed range regardless of the contaminant concentration detected. If the conce-