3 Operation

Model 6600

3-26 Part I

Teledyne Analytical Instruments

tration exceeds the upper limit of the range, the DC output will saturate at 1 V dc
(20 mA at the current output).

However, the digital readout and the RS-232 output of the concentration

are unaffected by the fixed range. They continue to read beyond the full-scale
setting until amplifier saturation is reached. Below amplifier saturation, the
overrange readings are accurate UNLESS the application uses linearization over
the selected range.

The concentration ranges can be redefined using the Range function

Manual screen, and the application gases can be redefined using the System
function, if they are not already defined as necessary.

CAUTION: Redefining applications or ranges might require

relinearization and/or recalibration.

To setup automatic ranging:

Enter the

Range

function to start the Range function.

Select range
mode : AUTO

If above screen displays MAN, use the

UP/DOWN switch to Select

AUTO, and Enter.

Press Escape to return to the previous Analyze Function.

3.6.3 Precautions

The Model 6600 allows a great deal of flexibility in choosing ranges for

automatic range switching. However, there are some pitfalls that are to be
avoided.

Ranges that work well together are:

Ranges that have the same lower limits but upper limits that differ
by approximately an order of magnitude

Ranges whose upper limits coincide with the lower limits of the
next higher range

Ranges where there is a gap between the upper limit of the range
and the lower limit of the next higher range.

Range schemes that are to be avoided include:

Ranges that overlap