Symptom

Probable Cause

Recommended
Corrective Action

Machine guards
and low oil mes-
sages are reversed

Improper wiring

Check the wiring to the
AMI’s J2 connector. The
machine guard switch
should be wired to pins 1
and 2, and the auto lube
system should be wired
to pins 3 and 4 (refer to
Input Devices (J2)).

Table limit switch-
es do not work

Improper wiring

Verify that the wires are
connected to pins 5 and 6
on the AMI’s J4 connec-
tor (refer to Emergency
Stop Inputs
).

Emergency stop
message is con-
stantly displayed
on

MILLPWR

oper-

ator console.

No External E-Stop is
wired to the J4 connector,
or it may be done so
incorrectly.

This input must be con-
figured. Either connect
an external emergency
stop switch to J4, or use
a wire jumper to defeat
it.