VEGA VEGAPULS 56 Profibus PA User Manual

Page 51

Advertising
background image

VEGAPULS 56 Profibus PA

51

Setup

The sensor electronics has two characteris-
tics points (at min. and max.) from which a
linear proportionality between product dis-
tance and the percentage of filling of the
vessel is generated.
Of course, the characteristics points must not
necessarily be at 0 % and 100 %, however
they should be as far apart as possible (e.g.
at 20 % and at 80 %). The difference be-
tween the characteristics points for the min./
max. adjustment should be at least 20 mm
product distance. If the characteristics points
are too close together, the possible measur-
ing error increases. Ideally, the adjustment
would be to carried out, as shown in the
example, at 0 % and at 100 %.

In the menu "

Instrument data/Parameter ad-

justment/Conditioning/Linearisation“ you can
enter later, if necessary, a correlation be-
tween product distance and % extent of filling
other than linear (see later subitem
Linearisation).

Scaling

• Click in the menu window "

Instrument data

parameter adjustment“ to "Conditioning

Conditioning

Conditioning

Conditioning

Conditioning“.

The menu window "

Conditioning“ opens.

• Click to "

Scaling

Scaling

Scaling

Scaling

Scaling“.

The displayed message tells you that you will
find this function on Profibus sensors under
the menu item "

Outputs“ in the menu window

"

Instrument data parameter adjustment“.

• Confirm the message with "

OK

OK

OK

OK

OK“.

• Click in the menu window "

Conditioning“ to

"

Quit

Quit

Quit

Quit

Quit“.

You are again in the menu window "

Instru-

ment data parameter adjustment“.

Scaling of the output signal

• Click in the menu window "

Instrument data

parameter adjustment“ to "Outputs

Outputs

Outputs

Outputs

Outputs“.

Advertising