VEGA VEGAPULS 56 Profibus PA User Manual

Page 82

Advertising
background image

82

VEGAPULS 56 Profibus PA

Technical data

Flange DIN DN 250
Material: 1.4571
Groove and tongue acc.
to DIN 2512 Form F, N

PN 16

PN 25

PN 40

PN 64

bar

40

25

16

64

-40

0

50

100

150

200

250

300

350

˚C

PN 16

PN 25

PN 40

bar

40

25

16

-40

0

50

100

150

200

250

300

350

˚C

Flange DIN DN 250
Material: 1.4571
Seal surface acc. to DIN
2526 Form B, C, D, E

Flanges acc. to ANSI (ASA) B16.5 seal surface RF, material 1.4571 in sizes 2“ to 10“ can be
used along the entire temperature range of -40°C … 350°C with the appropriate nominal
pressures from 150 lbs, 300 lbs, 600 lbs and 900 lbs.

Other flanges and appropriate process data on request.

Connection cables

Two-wire sensors

power supply and signal via one two-wire
cable

Electrical connection

- cable entry

for Aluminium and plastic housing:
one cable entry (four-wire: two- cable entries)
and spring-loaded terminal connection
up to max. 2.5 mm

2

wire cross-section.

- plug connection

optional for plastic housing:
four-pole, polarity reversal-safe screw
connection (four-wire: two plug connections)

Cable entry

- ia terminal compartment

1 … 2 x M20 x 1.5 (cable ø 5 … 9 mm)

Ground connection

max. 4 mm

2

Advertising