Product description

Adjustment with the adjustment pro-
gram VVO - VEGA Visual Operating

The setup and adjustment of the radar sen-
sors is generally done on the PC with the
adjustment program VEGA Visual Operating
(VVO) or with PACTware

TM

under Windows

®

.

The programs lead quickly through adjust-
ment and parameter setting by means of
pictures, graphics and process
visualisations.

Note:

The adjustment program VVO must be avail-
able in version 2.70 or higher.

The VEGA adjustment software VVO (VEGA
Visual Operating) operates either as a
subprogram of the host program PACTware

TM

acc. to the FDT concept (Field Device Tool) or
as an independent adjustment program on
any PC, engineering station or process con-
trol computer.

The adjustment program recognises the sensor type

Visualised input of a vessel linearisation curve in the
adjustment program VVO

VEGA’s adjustment program VVO can ac-
cess the adjustment options of VEGA sen-
sors in their entirety and, if necessary, can
update the complete sensor software. To do
this, the adjustment program must be in-
stalled on a PC which is equipped with a
Profibus Master Class 2 interface card or the
interface adapter VEGACONNECT 3.

The PC with the Profibus interface card can
be connected directly to any point on the DP
bus with the standard RS 485 Profibus cable.

In conjunction with the adapter
VEGACONNECT, the PC can be connected
directly to the sensor. VEGACONNECT com-
municates via a small jack directly with the
respective sensor.

The adjustment and parameter data can be
saved at any time on the PC with the adjust-
ment software and can be protected by
passwords. If necessary, the adjustments
can be transferred quickly to other sensors.