VEGASON 51P … 53P
Setup
The sensor electronics has two characteris-
tics points (at min. and max.) from which a
linear proportionality between product dis-
tance and the percentage of filling of the
vessel is generated.
Of course, the characteristics points must not
necessarily be at 0 % and 100 %, however
they should be as far apart as possible (e.g.
at 20 % and at 80 %). The difference be-
tween the characteristics points for the min./
max. adjustment should be at least 20 mm
product distance. If the characteristics points
are too close together, the possible measur-
ing error increases. Ideal would be to carry
out the adjustment as shown in the example,
at 0 % and at 100 %.
In the menu "
Instrument data/Parameter ad-
justment/Conditioning/Linearisation“ you can
enter later, if necessary, a correlation be-
tween product distance and % extent of filling
other than linear (see later subitem Linearisa-
tion).
Scaling
• Click in the menu window "
Instrument data
parameter adjustment“ to "
Conditioning“.
The menu window "
Conditioning“ opens.
• Click to "
Scaling“.
The message is displayed that you will find
this function on Profibus sensors under the
menu item "
Outputs“ in the menu window
"
Instrument data parameter adjustment“.
• Confirm the message with "
OK“.
• Click in the menu window "
Conditioning“ to
"
Quit“.
You are again in the menu window "
Instru-
ment data parameter adjustment“.
Scaling of the output signal
• Click in the menu window "
Instrument data
parameter adjustment“ to "
Outputs“.