VEGA VEGASON 51P…53P Profibus PA User Manual

Page 59

Advertising
background image

VEGASON 51P … 53P

59

Setup

• Click to "

Calculate“.

After a short calculation time, the levels as a
percentage of span and the corresponding
volume percentages are shown. The
outputted curve shows this correlation in a
diagram.

• Quit the linearisation table with "

OK“.

You are again in the menu window "

Tank

calculation“.

• Click to "

OK“ to save the tank calculation.

You are again in the menu window "

Linearisa-

tion -- user programmable curve --“. The
volume percentages, with the corresponding
level percentages, are shown. When clicking
in the bottom left part of the menu window to
"

Show scaled values“, liters will be displayed

according to the adjustment in the menu
"

Instrument data/Parameter adjustment/Con-

ditioning/Scaling“.

Calculate cylindrical tank

• Click in the menu window "

Linearisation --

user programmable curve --“ to
"

Calculate“ and in the menu window "Tank

calculation“ to the symbol for cylindrical
tanks.

Advertising