VEGASON 51P

53P

Adjust-
ment

Para-
meter

w.o
medium

Adjust-
ment
in

(min. adjustment)

OK

OK

+

OK

OK

OK

Ad-
just-
ment
in

at

OK

+

or

Max.

Min.

100 % (1.270 m) correspond
to 1200 liters

Span (4.58 m)

0 % (5.850 m) corresponds
to 45 liters

Adjustment is only necessary if the measure-
ment is made in gases deviating from air
(CO

2

, He, etc.). When measuring in gases,

sound the distance of the sensor to the prod-
uct surface and enter this in the menu item
"Measurement in gases“. The sensor can
then take the modified (compared to air)
sonic velocity in gases into account and
output correct levels.

Without special adjustment, the operating
range corresponds to the measuring range.
Generally, it is useful to choose a slightly
wider range (approx. 5 %) for the operating
range than for the measuring range.

Example:
Min./max. adjustment: 1.270 … 5.850 m;
adjust operating range to approx.
1.000 … 6.000 m.

Setup

Under the menu item "

Adjustment“ you inform

the sensor about the measuring range.

You can carry out the adjustment with or
without medium. Generally you will carry out
the adjustment without medium, as you can
then adjust without a filling/emptying cycle.

(adjustment independent of the level)

Key adjustment

Display indication

The distance indication flashes
and you can choose "feet“ and
"m“.

Confirm the adjustment with
"

OK“.

With "

+“ and "–“ you adjust the

percentage value for the min.
value (example 0.0 %).

The adjusted percentage
value is written in the sensor
and the distance for the min.
value corresponding to the
percentage value flashes