6 installation fault – VEGA VEGASON 71 - D User Manual

Page 10

Advertising
background image

VEGASON 71 - D

10

2.6

Installation fault

2 Technical Information

Fault:

Min. distance from sensor to h

max

not

maintained.

Problem: The upper range cannot be measured.
Solution: Correct min. distance of the sensor to 0,3 m.

Fault:

Sensor is not directed vertically to the surface.

Problem: Sensor receives weak echoes.
Solution: Mount sensor vertically.

Fault:

Strong heat fluctuations, e.g. sun.

Problem: The measuring accuracy is not constant.
Solution: Install a sun shield in the respective position.

Fault:

Surface has foaming problem.

Problem: The surface of the foam is detected as level.

The measured value is wrong.

Solution: The used measuring principle is not suitable

and must be replaced e.g. by a hydrostatic
system (pressure transmitter).

Fault:

Distance from sensor to h

max

too large.

Problem: Unreliable results due to very weak echoes.
Solution: Correct min. distance of the sensor to 0,3 m.

Fault:

Sensor too close to the flume wall, i.e. not

mounted in the centre.

Problem: Build-up and rough flume walls cause

measuring problems.

Solution: Mount sensor in the centre of the flume.

> 300 mm

h

max

< 300 mm

h

max

Advertising