7 electrical connection – VEGA VEGASON 71 - D User Manual

Page 11

Advertising
background image

VEGASON 71 - D

11

2.7

Electrical connection

2 Technical Information

2

1

3

4

5

9 10 11 12 13 14

21 22 23

15 16 17 18 19 20

24

+ - + -

6

7

8

1
2

1 2

1 2

+

-

1 2

Sensor

standard coax cable
max. cable length = 300 mm

Failure relay
output

Service switch

Power supply

Note
During operation the sensor is clocked with high-
performance impulses. It is therefore recommended to
provide all electrical connections first and then switch on
the power supply.

Fuse type TR5
Manufacturer, e.g. Messrs. Wickmann,
Current value see 2.2 Technical Data

Flow module:
-

Flow
Terminal 9 and 10

-

Sampling
Terminal 12 … 14

Level module

Current module

The connection line to the sensor can be extended as
shown. Therefore observe that a max. line length of 300 m
is not be exceeded.

standard waterproof
distributor box

Flow module

Current output

Relay output

Relay output

Service socket

Current module:
-

flow proportional
Terminal 21 and 22

-

level proportional
Terminal 23 and 24

Attention!

High

voltage