10.1 Error codes

Problemes during the measurement of wrong
adjustmentes during the programming are commented via
the displays with fault signals.

The fault signals are made in the mode "Indication of
measured values" via display 1 (top).

In below error schedule the fault signals sometimes cause
the fail safe relay to de-energize and the failure LED to
light.

The current outputs as well as the relay outputs react acc.
to the below schedule or their parameter adjustment in
mode

0 - 93

(see page 33).

When the error source is removed (e.g. repair of a line
break) the fault signal is cancelled.

Example of a fault signal:

1

E 3 . 0 4

Line break / temperature sensor defect

=

Coding of the error

=

Error class

=

E = Error

The error

E3.04

is indicated undelayed when switching on

and during operation after approx. 15 secs.

10.2 Error schedule

Error

Meaning

Pre-adjustment

Fail safe relay /

Relay output /

code

Mode

1 - 24

-LED

Current output

E2.01

Signal to noise distance

=

0

acc. to parameter adjustment

< 1 dB

in mode

0- 93

=

1

acc. to parameter adjustment
in mode

0 - 93

E2.02

Echo in dead time

without influence

Modification corresponds to

E2.03

Postoscillation > 1,8 m

the value for 100 %

E3.01

Data loss in the

without influence

E3.02

data memory

acc. to parameter adjustment

E3.03

in mode

0 - 93

E3.04

Line break/

=

0

acc. to parameter adjustment

Temperature sensor defect

in mode

0 - 93

=

1

acc. to parameter adjustment
in mode

0 - 93

ERR

Parameter exceeded

without influence

no change

during adjustment