VEGA VEGASON 71 - D User Manual
Page 21
VEGASON 71 - D
21
6 Flow measurement
Example of a linearization curve
0
50
100 %
50
%
100
4.L.05
100.0
4.L.01
010.0
4.H.01
030.0
4.H.05
100.0
Display
Flow
Level
Example of a linearity protocol
Index marker
Level percent
Flow percent
no.
(level)
01
30.0 %
10.0 %
02
50.0 %
15.0 %
03
75.0 %
40.0 %
04
90.0 %
70.0 %
05
100.0 %
100.0 %
L
= indication for flow percent
01
= 1. index markers
MODEFIELD
PARAMETER FIELD
flow percent value (factory setting)
Display 1
Display 2
4
= no. of the linearization curve
H
= indication for level percent
01
= 1. index markers
MODEFIELD
PARAMETER FIELD
level percent value (factory setting)
Display 1
Display 2
4
= no. of the linearization curve
4 . H
H
H
H
H . 0 1
1 0 0 . 0
Change-over
scrolling with key –>
scrolling with key + or –
4 . H
H
H
H
H . 0 1
4 . L
L
L
L
L . 0 1
4 . H . 0
0
0
0
0 1
4 . L . 9
9
9
9
9 1
4 . H . 0 1
11
11
4 . L . 0 9
9
9
9
9
4 . L
L
L
L
L . 0 1
1 0 0 . 0