5adjustment, 2 demonstration and programming example – VEGA VEGASON 71 - D User Manual
Page 18
Advertising
VEGASON 71 - D
18
5 Adjustment
5
Adjustment
5.1
Empty / full adjustment in metres without flow change
With this adjustment procedure two distances in m are defined which correspond to the levels of 0 % and 100 %.
5.2
Demonstration and programming example
Demonstration
0.000
0.350
0.700 0
100
m %
0.7 m distance ^ 0 %
0.35 m distance ^ 100 %
Programming example
0.350 m distance ^ 100 %
• Enquire mode
0 - 12
in the MODEFIELD
• Program
0.350
m in the PARAMETER FIELD
• Then store with key STO
0.700 m distance ^ 0 %
• Enquire mode
0 - 11
in the MODEFIELD
• Program
0.700
m in the PARAMETER FIELD
• Then store with key STO
(The sequence of the empty and full adjustment is individual)