4set-up, 1 flow chart for set-up – VEGA VEGASON 71 - D User Manual

Page 14

Advertising
background image

VEGASON 71 - D

14

4 Set-up

4

Set-up

4.1

Flow chart for set-up

Connect sensor to control electronics and
switch on the power supply.

The software version is displayed and the
fault signal responds.

The measuring system is ready and starts
the self-check cycle. Preliminary indication:
0.000 m

Self-check cycle, i.e.:
1. Echo limitation from top to bottom.
2. Echo limitation from bottom to top.
3. Generation of a measuring window, within

which a preferred processing of the echo
is made.

After approx. 1 … 3 min. the measuring
system is in operating status. The fault signal
extinguishes. The measuring system has
finished the self-check and operates with the
parameters of the factory setting.

?

Flow measurement

(level measurement in flume)

!

!

1

2

5

6

4

!

!

1

2

5

6

4

Optimization required.

In the flume a distance of e.g.
0,600 m is determined.

NO

YES

48.94

0.600

m

0.000

m

Feeding

phase:

1.

2.

3.

Sensor

Echo

Measuring
window

Flume