Luminex xPONENT 3.1 Rev 2 User Manual

Page 45

Advertising
background image

% CV of microspheres - The measure of relative dispersion within the distribution.

%CV = 100 x Std Dev / Mean

Standard Deviation - For calculating sample variability or dispersion, Luminex uses the

standard deviation formula.

Peak - The value that is equal to the largest number of data points within the distribution.

For example, in data set {1,2,2,3,3,3,4,5}, 3 is the peak because it occurs the largest
number of times in the distribution list.

Trimmed Count*

Trimmed Mean*

Trimmed%CV of microspheres*

Trimmed Standard Deviation*

Trimmed Peak*

% CV of Replicates - The measure of relative dispersion within the distribution of results

for replicate samples.

%CV = 100 X Std Dev / Mean

% Recovery - A measure of how accurately your observed results match your expected

results following regression analysis.

(Observed concentration) / (Expected concentration) x 100%

Expected Result - The known or expected test result value for a standard or control.

Control Range - Low - The lowest value for an assay control used to determine pass/fail

criteria for an assay.

Control Range - High - The highest value for an assay control used to determine pass/fail

criteria for an assay.

Normalized Net Median - For each analyte in a well the Normalized Net Median (NNM) =

(net median of analyte) / (net median of normalization bead)

Units - The unit of measure for an analyte, for example, pg/mL.

*Trimmed statistics remove the lower and upper five percent of the extreme statistical values,
then use the remaining values for the Mean, Standard Deviation, or %CV calculations.

The statistics displayed change according to the analysis type selected.

Analyte - Contains a list of analytes run in the batch. Select an analyte to view all statistics

for that analyte.

Current Well - Displays the statistics of the well currently being displayed (This changes to

Displayed Well if viewing a batch using the Open button of the Saved Batches tab).

Single Step - Allows analysis one well at a time. Click to turn function on or off.

Results - Displays statistics related to the batch. Use the up, down, left, and right arrow

buttons to move through the table, or use the scroll bars. The maximize/minimize toggle
button expands the batch table and returns it to the standard size.

Plate - A list of the available plates (if more than one).

WARNING:

If using multiple plates, be careful to ensure that plates are used
in the proper order. Failure to do so can result in inaccurate data
and test results.

Histogram/Dot Plot - Displays real-time monitoring of batch sampling during acquisition.