Multichannel Systems Cardio2D Manual User Manual

Page 52

Advertising
background image

Cardio2D Manual

48

The drop down menu "Min/Max Type" describes the type of the minimum or maximum "Local"
or "Global". These options refer to one single field potential signal: "Local" means the first
detected minimum or maximum, "global" means the deepest minimum or highest maximum
in the field potential within the Min./Max. search interval. Please see the picture.

Whether you select the local or the global minimum or maximum, depends on the aim of the user:
If you are interested in the maximal slope of the action potential, you have to select the first local
minimum or maximum. If you are interested in absolute minimal or maximal values, you have to
select the global minimum or maximum.

Please select the time used for searching the minimum or maximum of the field potential
"Min/Max Search" from the drop down menu (range from +/- 1 ms up to +/- 200 ms).

You can set a "HB timestamp" to each detected field potential either to the maximum or to the
minimum or to the maximal slope of the field potential. This timestamp is visualized in the plot
with a black dot, and stored in the raw data file.

Please read also chapter "Detection Parameter".

Click "HB Average" tab in Analyzer Settings.

It is possible to display several recordings of an electrode channel in overlay and to average the
values. In this dialog the pre and post time before and after a heartbeat can be selected via up
down box. Enable the check box "On" to display and save the averaged data traces. If you do not
need the average function, disable the check box to save computer performance and disk space.