3 mea window cardio2d – Multichannel Systems Cardio2D Manual User Manual

Page 88

Advertising
background image

Cardio2D Manual

84

In this screenshot a MEA electrode grid 8 x 8 is displayed. Different layout grids are possible,
depending on the selected raw data file. The numbers of the electrodes refer to their position
in the grid, to rows and columns. Using a different type of a MEA, for example, the layout will
be adapted.

If you have different clusters of cells with different pacemakers on the MEA, you have to analyze
each cluster separate. Mark the electrodes belonging to the first cluster by using the "Add

Channel" icon

to select channels. Deselect the other electrodes by clicking the "Remove

Channels" icon

. Run the analysis and save the results. After doing this, you can mark the next

cluster for running a second analysis.

Settings File

In the "Settings file" all settings of the currently analyzed Cardio2D result or raw data file are
displayed in a tree view. The window contains all information saved during recording the file,
and you can read any information you need for analyzing the file. It is not useful, and therefore
not possible to change these information in a Cardio2D raw data file. If you re-analyze a raw data
file with new parameters you can save the new results in a new Cardio2D result file.

6.3

MEA Window Cardio2D+

The MEA window on the right side is divided into separate window panes, and has five tabs:
"Field Potential", "ROI Selection", "Heartbeat", "HB Averager" and "QT-related Analysis".
Switching between the tabs influences the settings of each of the separate windows. Select
the single channels of each tab via "Selection control" window of the "Control Window"
of Cardio2D+.