Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide

159

Programming the Status Register System

Status Groups

Standard Operation Status Group

NOTE

Some of the bits in this status group do not apply to the E4428C, E8257D, E8267D, E8663B,
and the N5181A/82A/83A, and returns zero when queried. See

Table 4- 5 on page 160

for

more information.

The Agilent MXG SCPI command :STAT:OPER:SUPP, can suppress the managing of this
status group and save 50 us from the switching time. Refer to the SCPI Command Reference.

The Operation Status Group is used to determine the specific event that set bit 7 in the

Status Byte

Register

. This group consists of the

Standard Operation Condition Register

, the

Standard Operation

Transition Filters (negative and positive)

, the

Standard Operation Event Register

, and the

Standard

Operation Event Enable Register

.

This is the named
status register for
the E4438C.
However, not all
signal generator
models use all of
the shown events
(i.e. some use only
a subset of the
E4438C’s status
registers).