Creating and Downloading User-Data Files
Data Transfer Troubleshooting (N5182A, E4438C and E8267D Only)

Data Requirements

The signal generator requires a file with a minimum of 60 bytes

For every data bit (bit 0), you must provide 7 bits of control information (bits 1- 7).

ERROR - 223, Too much data

PRAM download exceeds the size of PRAM memory.

Either use a smaller pattern or get more memory by ordering the appropriate
hardware option.

Table 8-14 PRAM Data Byte

Bit

Function

Value

Comments

0

Data

0/1

This is the data (payload) bit. It is “unspecified” when burst (bit 2) is set to 0.

1

Reserved

0

Always 0

2

Burst

0/1

1 = RF on
0 = RF off
For non- bursted, non- TDMA systems, to have a continuous signal, set this bit to 1 for all
bytes. For framed data, set this bit to 1 for on timeslots and 0 for off timeslots.

3

Reserved

0

Always 0

4

Reserved

1

Always 1

5

Reserved

0

Always 0

6

EVENT1
Output

0/1

To have the signal generator output a single pulse at the EVENT 1 connector, set this bit
to 1. Use this output for functions such as a triggering external hardware to indicate when
the data pattern begins and restarts, or creating a data- synchronous pulse train by
toggling this bit in alternate bytes.

7

Pattern Reset

0/1

0 = continue to next sequential memory address.
1 = end of memory and restart memory playback.
This bit is set to 0 for all bytes except the last byte of PRAM. To restart the pattern, set
the last byte of PRAM to 1.

Table 8-13 PRAM Download - Symptoms and Causes

Symptom

Possible Cause