Agilent Technologies N5183A MXG User Manual

Page 46

Advertising
background image

36

Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide

Using IO Interfaces
Using LAN

NOTE

The following sections are specific to Agilent IO Libraries versions M and earlier and apply
only to the Windows NT platform.

Using VISA Assistant

Use the VISA Assistant, available with the Agilent IO Library versions M and earlier, to communicate
with the signal generator over the LAN interface. However, you must manually configure the VISA
LAN client. Refer to the Help menu for instructions on configuring and running the VISA Assistant
program.

1.

Run the IO Config program.

2.

Click on TCPIP0 in the Available Interface Types text box.

3.

Click the

Configure button. Then Click OK to use the default settings.

4.

Click on TCPIP0 in the Configured Interfaces text box.

5.

Click

Edit...

6.

Click the

Edit VISA Config... button.

7.

Click the

Add device button.

8.

Enter the TCPIP address of the signal generator. Leave the Device text box empty.

9.

Click the

OK button in this form and all subsequent forms to exit the IO Config program.

If You Have Problems

1.

Verify the signal generator’s IP address is valid and that no other instrument is using the IP
address.

2.

Switch between manual LAN configuration and DHCP using the front- panel

LAN Config softkey and

run the ping program using the different IP addresses.

NOTE

For Agilent IO Libraries versions M and earlier, you must manually configure the VISA LAN
client in the IO Config program if you want to use the VISA Assistant to verify LAN
configuration. Refer to the IO Libraries Installation Guide for information on configuring IO
interfaces. The IO Config program interface is shown in

Figure 2- 4 on page 38

.