Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide

47

Using IO Interfaces

Using RS-232 (ESG, PSG, and E8663B Only)

2.

Setting the RS- 232 Echo Softkey

3.

Connect an RS- 232 cable from the computer’s serial connector to the ESG signal generator’s
RS-232 connector or the PSG or E8663B’s AUXILIARY INTERFACE connector. Refer to

Table 2- 2

for

RS- 232 cable information.

NOTE

Any 9 pin (male) to 9 pin (female) straight- through cable that directly wires pins 2, 3, 5, 7,
and 8 may be used.

Table 2-2

RS-232 Serial Interface Cable

Quantity

Description

Agilent Part Number

1

Serial RS- 232 cable 9- pin (male) to 9- pin (female)

8120- 6188

For details on each key, use the key help (described in User’s Guide). For additional SCPI command information, refer to the SCPI
Command Reference.

Toggle RS-232 Echo Off On until Off is
highlighted. Selecting On echoes or returns
characters sent to the signal generator and
prints them to the display.

SCPI commands:

:SYSTem:COMMunicate:SERial:ECHO ON|OFF
:SYSTem:COMMunicate:SERial:ECHO?