Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide

179

Programming the Status Register System

Status Groups

Data Questionable Calibration Event Enable Register

The Data Questionable Calibration Event Enable Register lets you choose which bits in the Data
Questionable Calibration Event Register set the summary bit (bit 8 of the Data Questionable
Condition register) to 1.

Command:

STATus:QUEStionable:CALibration:ENABle <value>

, where <value> is the sum of the decimal

values of the bits you want to enable.

Example:

Enable bit 1 and bit 0 so that whenever either of those bits are set to 1, the Data Questionable
Calibration summary bit of the Data Questionable Condition Register is set to 1. Send the command
STAT:QUES:CAL:ENAB 520

(2 + 1).

Query:

STATus:QUEStionable:CALibration:ENABle?

Response:

Decimal value of the sum of the bits previously enabled with the
STATus:QUEStionable:CALibration:ENABle <value>

command.