Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide

197

Creating and Downloading Waveform Files

Waveform Phase Continuity

The following figures illustrate the influence a single sample can have. The generated 3- tone test
signal requires 100 samples in the waveform to maintain periodicity for all three tones. The
measurement on the left shows the effect of using the first 99 samples rather than all 100 samples.
Notice all the distortion products (at levels up to

−35 dBc) introduced in addition to the wanted

3- tone signal. The measurement on the right shows the same waveform using all 100 samples to
maintain periodicity and avoid a phase discontinuity. Maintaining periodicity removes the distortion
products.

Sampled Sinewave with No Discontinuity

Waveform length

Added sample

3-tone - 20 MHz Bandwidth

3-tone - 20 MHz Bandwidth

Measured distortion = 35 dBc

Phase Continuity

Measured distortion = 86 dBc

Phase Discontinuity