164

Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide

Programming the Status Register System
Status Groups

Baseband Operation Event Register

The Baseband Operation Event Register latches transition events from the condition register as
specified by the transition filters. Event registers are destructive read only. Reading data from an
event register clears the content of that register.

Baseband Operation Event Enable Register

The Baseband Operation Event Enable Register lets you choose which bits in the Baseband Operation
Event Register can set the summary bit (bit 7 of the Status Byte Register).

Query:

STATus:OPERation:BASeband[:EVENt]?

Command:

STATus:OPERation:BASeband:ENABle <value>

, where

<value>

is the sum of the decimal values of the bits you want to enable.

Example:

Enable bit 0 and bit 1 so that whenever either of those bits are set to 1, the Baseband Operation
Status summary bit of the Status Byte Register is set to 1. Send the command STAT:OPER:ENAB (2 +
1).

Query:

STATus:OPERation:BASeband:ENABle?

Response:

Decimal value of the sum of the bits previously enabled with the
STATus:OPERation:BASeband:ENABle <value>

command.