Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide

Getting Started with Remote Operation
Using the Web Browser

LAN Configuration System Defaults (Agilent MXG)

NOTE

The instrument’s LAN configuration system information can be found on the signal
generator’s homepage and on the signal generator. Refer to

“Enabling the Signal Generator

Web Server” on page 11

and to

“Displaying the LAN Configuration Summary (Agilent MXG)”

on page 15

.

If the instrument has been restored to the factory defaults from the LAN Setup menu the signal
generator will revert to the values displayed in

Table 1- 1 on page 14

. Refer to

“Displaying the LAN

Configuration Summary (Agilent MXG)” on page 15

.

To reset the instrument LXI password to “agilent” and the LAN settings to their factory default
values, press the following key sequence on the signal generator:

Utility > I/O Config > LAN Setup > Advanced Settings > Restore LAN Settings to Default Values >
Restore LAN Settings to Default Values

NOTE

There are no SCPI commands associated with this LXI password factory reset.

For more information, refer to the signal generator’s Web Server Interface Help.

Table 1-1

LAN Configuration Summary Values

Parameter

Default

Signal Generator LAN Configuration Summary

Hostname:

Agilent–<model number>–<last_5_chars_of_serial_number>

Config Type:

AUTO

IP Address:

127.0.0.1

Connection Monitoring:

On

Subnet:

255.255.255.0

DNS Server Override:

Off

Gateway:

0.0.0.0

Dynamic DNS Naming:

On

RFC NETBIOS Naming:

On

DNS Server:

0.0.0.0