NOTE: “Discharge” will normally use the Setpoint 3 output. If Setpoint 3 is

unavailable, discharge will use the Setpoint 4 output

8. If Setpoint 4 is unused, the display will show “Zb x” where x = 0 or 1. Enter 1

to enable or 0 to disable Zeroband.

9. If “SP bA” , “SP AbA” or “SP Abr” was selected, the display will show

“SPi xxx”, where “xxx” is OFF, LO or HI. Enter OFF to disable the setpoint
software inhibit, LO to enable the software inhibit when the gross weight is
less than Setpoint 5, or HI to enable the software inhibit when the gross
weight is greater than or equal to Setpoint 5.

NOTE: Unlike Setpoint 1 through 4, Setpoint 5 does not have an output. The Setpoint 5

value is the Setpoint Software Inhibit reference. A batch, auto batch, or autobatch with

repeat cycle will be inhibited from starting if the weight on the scale is less than or

greater than (as defined by the user) the Setpoint 5 value.

10. If “SP AbA” or “SP Abr” is selected, the display will show “SPA2 x” where x

equals 0 or 1. Enter 1 to enable the Setpoint Autozero feature or 0 to
disable. If enabled, the instrument will automatically perform a zero
operation before starting an auto batch cycle. Enable if the application uses
a platform, disable if the application uses a hopper.

11. If “SP AbA” (Auto Batch mode) or “SP Abr” (Auto Batch with Repeat mode)

was selected, the instrument will now prompt the user for the amount of
time to delay between each ingredient, “d1” (DELAY 1), before starting the
next batch.
Enter the amount of time (from 0-99 seconds).

12. In all but the checkweigh setpoint mode, the display will now show “d2 xx”,

where “xx” is the delay 2 in seconds. The discharge output will turn off 2
seconds after the cause of any software inhibit is removed. If delay 2 is set
to 0, the discharge output will turnoff when the STOP key is pressed. Enter
the desired amount of delay, from 0 to 99 seconds.

NOTE: If 0 is selected, the instrument will wait for the next setpoint to be manually

started by the operator pressing the START ‘’ button.

13. If “SP Aba” or “SP Abr” was selected, then the display will show “print x”,

where x is a 0 or 1.

0 = No print function is enabled.
1 = Print function enabled and the printer will automatically print at the
end of the delay period.

14. The unit will return to the normal weight display.