OPEN (OFF)

CLOSED (ON)

Switch Settings DSW1 =

1,2,4,5,6,8

3,7

DSW2 = 1,2,3,4,5,6,7,8

NONE

9600 baud, 8 characters, non parity, 1 stop bit, cts=0

8. 50-3960 Ticket Printer Setup

50-3960 Switch Settings

SW1 =

12345678910
1110000000

9600, None, 8 Bits, Busy=0

Use cable Acc 1264 (3-93081-1)

9. PTR-2242 / 2642 / 2742 / 2844 = "P 2642" Setup:

9600 Baud
No Parity
8 Bits
1 Stop bit

Use Cable Acc-1269 ( 3-93212-2)

The printer must be connected to the 5200-A via cable ACC-1269 and have power ON
before selecting the "P-2642" device. Once selected, choose tdi options, nEt or Gtn
options, Cr/LF options then no bAr/bAr at which time the 5200-A will download data to
the PTR-2642. There will be a short period when the 5200-A will blank, then return to
the gross weigh mode.

General Notes regarding the 5200A's outputs:

1. It is recommended to use a baud rate higher than 300 on com port 2. The 300 baud

rate on that channel may affect the operation of the 5200-A.

2. If display (RMT-140X-A) ( 20mA) is on port 1, and a printer is interfaced to com port

2, the CTS setting of com port 1 must be set to 0.

3. If either the 50-3960 or PTR-3950 ticket printers are interfaced to the 5200-A

indicator, and Autoprint is enabled in either AbA or Abr setpoint modes, a ticket
must be inserted in the printer before the batch is started.

4. When setting up the 140X-A remote display, the remote may show "SUCCESS" and

still not communicate. Check and compare settings manually in both the indicator
and the remote display.

5. If SPon = 1, and using "SP" setpoint mode, any attached relays will be energized

until the setpoint limits are reached. A brief contact may be made when power-up
occurs or certain keys are pressed. Use caution when operating or testing.

Appendix IV: WIRING DIAGRAM for SERIAL OUTPUTS

50515

46

Issue #2 4/04