standard CTS operation or printer with busy, active low).

Serial Output Operation and Programming:

Clear to Send (CTS) Selection
Port 1 Only

CTS Value

50-3710

1

50-3715

1

50-3920 (Oki 182)

0

50-3921 (Oki 184T)

1

50-3925/SP2000/SP2200

0

PTR-3950/610

0

50-3960/295/590

0

Computer with CTS Handshake

1

Remote Display

0

20mA Current Loop

0

With ACC 146 Installed
Port 1

&

Port 2

CTS Value

50-3715

50-3921

1

50-3715

PTR-3950

1

50-3715

50-3960

1

SP-2000

1

50-3960

50-3715

0

50-3960

50-3921

0

50-3960

PTR-3950

0

50-3921

50-3715

1

50-3921

50-3960

1

50-3921

PTR-3950

1

PTR-3950

50-3715

0

PTR-3950

50-3921

0

PTR-3950

50-3960

0

Switch settings on ACC-146: 1, 3, 5, and 8 ON.

h.

The instrument will now return to the normal weigh mode.

2. Programming the Serial IO Format

There are 3 channels available

Channel 1 - Standard Serial Port
Channel 2 - RS232 / 20mA CL
Channel 3 - RS422/485 Output

a. Press FUNC, 3 to enter IO calibration mode. The display will show ‘CAL IO”.

Press the number of the IO channel you wish to format.

50515