Configuring hardware – Agilent Technologies E8247C PSG CW User Manual

Page 153

Advertising
background image

Chapter 6

143

Custom Arb Waveform Generator

Configuring Hardware

Configuring Hardware

“To Set the ARB Reference”

see

page 144

To Set a Delayed, Positive Polarity, External Single Trigger

Using this procedure, you learn how to utilize an external function generator to apply a delayed
single-trigger to a custom multicarrier waveform.

1. Connect an Agilent 33120A function generator or equivalent to the signal generator PATT TRIGGER IN

port, as shown in

Figure 6-1

.

Figure 6-1

2. On the signal generator, press

Preset

.

3. Press

Mode

>

Custom

>

Arb Waveform Generator

.

4. Press

Multicarrier Off On

until On is highlighted.

5. Press

Trigger

>

Single

.

6. Press

Trigger

>

Trigger Setup

>

Trigger Source

>

Ext

.

7. Press

Ext Polarity Neg Pos

until Pos is highlighted.

8. Press

Ext Delay Off On

until On is highlighted.

9. Press

Ext Delay Time

>

100

>

msec

.

The Custom Arb Waveform Generator has been configured to play a single multicarrier waveform
100 milliseconds after it detects a change in TTL state from low to high at the PATT TRIG IN rear panel
connector.

10. Set the function generator waveform to a 0.1 Hz square wave at an output level of 0 to 5V.

Advertising