Sweep trigger source – Agilent Technologies AGILENT U2761A User Manual

Page 101

Advertising
background image

Features and Functions

2

U2761A User’s Guide

83

Soft Front Panel Operation

On the Sweep Time panel as shown in

Figure 2- 28

, input

the desired sweep time and select the unit from the drop
down list.

Remote Interface Operation

SWEep:TIME <seconds>

Sweep Trigger Source

In the sweep mode, the U2761A outputs a single sweep
when a trigger signal is received. After one sweep from the
start frequency to the stop frequency, the U2761A will
continuously output the preset waveform at the start
frequency while waiting for the next trigger.

The sweep trigger source consists of Internal (immediate),

External, or Manual (software). The default trigger source
is Internal
.

When the Internal source is selected, the U2761A outputs

a continuous sweep at a rate determined by the specified
sweep time.

When the External source is selected, the U2761A accepts

a hardware trigger applied to the device Trig In
connector. The U2761A initiates one sweep each time
Trig In receives a TTL pulse with the specified polarity.

When the Manual source is selected, the U2761A outputs

one sweep each time the soft front panel

Trigger

is

clicked.

Soft Front Panel Operation

On the Trigger Setup panel as shown in

Figure 2- 28

, select

the desired sweep trigger source from the drop down list

at the Source panel.

Advertising