U2761A User’s Guide

2

Features and Functions

To specify whether the U2761A is triggered on the rising or
falling edge of the signal from the Trig In connector, select
Positive

for rising edge triggering and

Negative

for falling edge triggering.

Remote Interface Operation

TRIGger:SOURce {IMMediate|EXTernal|BUS}

Use the following command to specify whether the U2761A
is triggered on the rising or falling edge of the signal from
the Trig In connector.

TRIGger:SLOPe {POSitive|NEGative}

Set Frequency Sweep Using SCPI Commands

Example 1

-> *CLS

; *RST

//To reset the U2761A to default

power

-on state, this command

can be ignored if this operation
is not required.

-> FUNC SIN

//Sets the carrier waveform to

Sine wave.

-> VOLT 5 VPP

//Sets the output amplitude to

5 Vpp.

-> VOLT:OFFS 0

//Sets the output offset to 0.

-> SWE:STAT ON

//Enables frequency sweep.

-> FREQ:STAR 500

//Changes the start frequency to

500 Hz.

-> FREQ:STOP 10000

//Changes the stop frequency to

10 kHz.

-> SWE:SPAC LIN

//Changes the sweep mode to

Linear.

-> SWE:TIME 2

//Changes the sweep time to 2 s.

-> TRIG:SOUR EXT

//Changes the sweep trigger

source to External.

-> TRIG:SLOP POS

//Changes the U2761A to be

triggered on the rising edge.

-> OUTP ON

//Turns on output.

Advertising