Pulse width, Figure – Agilent Technologies AGILENT U2761A User Manual

Page 64

Advertising
background image

46

U2761A User’s Guide

2

Features and Functions

Pulse Width

The pulse width represents the time from the 50% threshold
of the pulse’s rising edge to the 50% threshold of the next
falling edge.

The range of the pulse width is 40 ns to <2000 s (see

restrictions below). The default pulse width is 500

μ

s.

The minimum pulse width (Wmin) is affected by the

period.

Wmin = 40 ns for period

≤ 10 s

Wmin = 200 ns for period > 10 s, but

≤ 100 s

Wmin = 2

μs for period > 100 s, but ≤ 1000 s

Wmin = 20

μs for period > 1000 s

The specified pulse width must also be less than the

difference between the period and the minimum pulse
width as shown in the equation below. The U2761A will
adjusts the pulse width as necessary to accommodate the
specified period.

Pulse Width

Period – Wmin

Soft Front Panel Operation

After selecting

Pulse

at the top panel as shown in

Figure 2- 2

, on the Width panel below, input the desired

pulse width value and select the unit from the drop down
list.

Figure 2-13 Panel view of the pulse width section

Remote Interface Operation

FUNCtion:PULSe:WIDTh <seconds>