WAVEFORMS

Standard

Sine, Square, Ramp, Triangle, Pulse, DC

Built-in arbitrary

Exponential Rise, Exponential Fall, Negative Ramp

WAVEFORM CHARACTERISTICS

Sine

Frequency range

1

μHz to 20 MHz (1 μHz resolution)

Amplitude flatness

1

(relative to 1 kHz)

<100 kHz

0.2 dB

100 kHz to 1 MHz

0.35 dB

1 MHz to 20 MHz

0.7 dB

Harmonic distortion

2

Frequency range

<1 Vpp

≥1 Vpp

DC to 20 kHz

–70 dBc

–60 dBc

20 kHz to 100 kHz

–65 dBc

–60 dBc

100 kHz to 1 MHz

–50 dBc

–45 dBc

1 MHz to 20 MHz

–40 dBc

–35 dBc

Total harmonic distortion

2

DC to 20 kHz

0.10%

Spurious (Non-harmonic) output

3

DC to 1 MHz

–65 dBc

1 MHz to 20 MHz

–65 dBc + 6 dB/octave

Phase noise (10 kHz offset)

–115 dBc/Hz (Typical)

Square

Frequency range

1

μHz to 20 MHz (1 μHz resolution)

Rise/fall time

<18 ns, 10 to 90% terminated load (50

Ω)

Overshoot

<2%

Variable duty cycle

20% to 80% (up to 10 MHz)
40% to 60% (up to 20 MHz)

Asymmetry (@ 50% duty)

1% of period + 5 ns

Jitter (RMS)

> 50 kHz = 1 ns + 100 ppm of period
≤ 50 kHz = 10 ns + 100 ppm of period

Advertising