Dc offset voltage – Agilent Technologies AGILENT U2761A User Manual

Page 51

Advertising
background image

Features and Functions

2

U2761A User’s Guide

33

Remote Interface Operation

VOLTage <amplitude>

Or, you can set the amplitude by specifying a high level and
low level using the following commands.

VOLTage:HIGH <voltage>

VOLTage:LOW <voltage>

You can also use the APPLy command to select the function,
frequency, amplitude, and offset.

DC Offset Voltage

The default offset is 0 V for all functions.

Limits Due to Amplitude

The relation between the offset voltage and output amplitude
is shown below. Vmax is the maximum peak voltage for the
selected output termination (5 V for a 50

Ω load or 10 V for

a high

-impedance load).

If the specified offset voltage is not valid, the U2761A will
automatically adjusts it to the maximum DC voltage allowed
with the specified amplitude.

V

offset

V

max

V

pp

2

---------