Output amplitude – Agilent Technologies AGILENT U2761A User Manual

Page 49

Advertising
background image

Features and Functions

2

U2761A User’s Guide

31

Remote Interface Operation

FREQuency <frequency>

You can also use the APPLy command to select the function,
frequency, amplitude, and offset.

Output Amplitude

The default amplitude is 1 Vpp (into 50

Ω

) for all functions.

Offset Voltage Limitations

The relation between the output amplitude and offset voltage
is shown below. Vmax is the maximum peak voltage for the
selected output termination (5 V for a 50

Ω load or 10 V for

a high

-impedance load).

Limits Due to Output Termination

If you change the output termination setting, the value of
the output amplitude will be adjusted (and no error will be
generated).

For example, if you set the amplitude to 5 Vpp and then
change the output termination from 50

Ω to “high

impedance”, the amplitude value will double to 10 Vpp. If
you change

from “high impedance” to 50

Ω, the displayed

amplitude value will drop to half. For more information, see

“Output Termination”

on page 36.