Figure 2-28 Panel view of sweep

Remote Interface Operation

SWEep:STATe {0|OFF|1|ON}

Start Frequency and Stop Frequency

The start frequency and stop frequency sets the upper and
lower frequency bounds for the sweep. The U2761A begins at
the start frequency, sweeps to the stop frequency, and then
resets back to the start frequency.

The range of the start and stop frequencies is 1

μHz to

20 MHz (limited to 200 kHz for Ramp and Arbitrary
waveforms). The sweep is phase

-continuous over the full

frequency range. The default start frequency is 100 Hz. The
default stop frequency is 1 kHz.

To sweep up in frequency, set the start frequency < stop

frequency. To sweep down in frequency, set the start
frequency > stop frequency.

Soft Front Panel Operation

On the Start Frequency and Stop Frequency panels as
shown in

Figure 2- 28

, input the desired start frequency and

stop frequency values and select the respective units from
the drop down lists.